International Catalog

ENDOCRINOLOGY - FERTILITY

LIAISON® LH

312201

Intended use

For the quantitative determination of human Luteinizing Hormone (hLH).

Material provided

Reagent Integral for 100 determinations.

Method

1-step immunoluminometric sandwich assay using directly coated magnetic microparticles

Detection limit

0.2 mIU/mL (WHO 1st IRP 80/552)

Calibrators

Included

Incubations

10 min

Measuring range

0.2-250 mIU/mL

Sample type

Serum/EDTA-plasma

Sample size

50 µL

LIAISON® FSH

312251

Intended use

For the quantitative determination of human Follicle-Stimulating Hormone (hFSH).

Material provided

Reagent Integral for 100 determinations.

Method

1-step immunoluminometric sandwich assay using directly coated magnetic microparticles

Detection limit

0.25 mIU/mL (WHO IRP 94/632)

Calibrators

Included

Incubations

10 min

Measuring range

0.25-400 mIU/mL

Sample type

Serum/citrated plasma

Sample size

25 µL

LIAISON® Prolactin

312151

Intended use

For the quantitative determination of human Prolactin (hPRL).

Material provided

Reagent Integral for 100 determinations.

Method

1-step immunoluminometric sandwich assay using directly coated magnetic microparticles

Detection limit

10 mIU/L (WHO 3rd IRP 84/500)

Calibrators

Included

Incubations

10 min

Measuring range

10-8000 mIU/L

Sample type

Serum/EDTA-plasma

Sample size

25 µL

LIAISON® Estradiol

310400

Intended use

For the quantitative determination of human estradiol

Material provided

Reagent Integral for 100 determinations

Method

Competitive 1-step immunoluminometric assay using directly coated magnetic microparticles

Detection limit

12 pg/mL

Calibrators

Included

Incubations

24 min

Measuring range

12-1100 pg/mL

Sample type

Serum/Plasma

Sample size

200 µL

LIAISON® Progesterone

310420

Intended use

For the quantitative determination of human progesterone

Material provided

Reagent Integral for 100 determinations

Method

Competitive 1-step immunoluminometric assay using directly coated magnetic microparticles

Detection limit

0.4 ng/mL

Calibrators

Included

Incubations

21 min

Measuring range

0.4-40 ng/mL

Sample type

Serum/Plasma

Sample size

75 µL

LIAISON® Testosterone

310410

Intended use

For the quantitative determination of Testosterone

Material provided

Reagent Integral for 100 determinations

Method

Competitive 1-step immunoluminometric assay using directly coated magnetic microparticles

Detection limit

0.05 ng/mL

Calibrators

Included

Incubations

12 min

Measuring range

0.05-16 ng/mL

Sample type

Serum/Plasma

Sample size

100 µL

LIAISON® HCG

312311

Intended use

For the quantitative determination of human Chorionic Gonadotropin (hCG and

Material provided

Reagent Integral for 100 determinations.

Method

1-step immunoluminometric sandwich assay using directly coated magnetic microparticles

Detection limit

0.5 mIU/mL (3rd IS 75/537)

Calibrators

Included

Incubations

10 min

Measuring range

0.5-5000 mIU/mL

Sample type

Serum

Sample size

30 µL